pH Range | 0.00 to 14.00 pH |
---|
pH Resolution | 0.01 pH |
---|
pH Accuracy | ±0.02 pH |
---|
pH Calibration | automatic one, two, or three points with automatic recognition of five standard buffers (pH 4.01, 6.86, 7.01, 9.18, 10.01) or one custom buffer |
---|
mV Range | ±600.0 mV |
---|
mV Resolution | 0.1 mV |
---|
mV Accuracy | ±0.5 mV |
---|
ORP Range | ±2000.0 mV |
---|
ORP Resolution | 0.1 mV |
---|
ORP Accuracy | ±1.0 mV |
---|
ORP Calibration | automatic at one custom point (relative mV) |
---|
EC Range | 0 to 9999 µS/cm; 0.000 to 200.0 mS/cm (absolute EC up to 400.0 mS/cm) |
---|
EC Resolution | auto-ranging: 1 µS/cm from 0 to 9999 µS/cm; 0.01 mS/cm from 10.00 to 99.99 mS/cm; 0.1 mS/cm from 100.0 to 400.0 mS/cm; auto-ranging (fixed mS/cm): 0.001 mS/cm from 0.000 to 9.999 mS/cm; 0.01 from 100.0 to 400.0 mS/cm ; manual: 1 µS/cm; 0.001 mS/cm; 0.01 |
---|
EC Accuracy | ±1% of reading or ±1 µS/cm whichever is greater |
---|
EC Calibration | automatic single point, with six standard solutions (84 µS/cm, 1413 µS/cm, 5.00 mS/cm, 12.88 mS/cm, 80.0 mS/cm, 111.8 mS/cm) or custom point |
---|
TDS Range | 0 to 9999 ppm (mg/L); 0.000 to 400.0 ppt (g/L) (maximum value depends on TDS factor) |
---|
TDS Resolution | manual: 1 ppm (mg/L); 0.001 ppt (g/L); 0.01 ppt (g/L); 0.1 ppt (g/L); 1 ppt (g/L); auto-ranging: 1 ppm (mg/L) from 0 to 9999 ppm (mg/L); 0.01 ppt (g/L) from 10.00 to 99.99 ppt (g/L); 0.1 ppt (g/L) from 100.0 to 400.0 ppt (g/L); auto-ranging (fixed ppt): |
---|
TDS Accuracy | ±1% of reading or ±1 ppm (mg/L), whichever is greater |
---|
TDS Calibration | based on conductivity calibration |
---|
Resistivity Range | 0 to 999999 Ω•cm; 0 to 1000.0 kΩ•cm; 0 to 1.0000 MΩ•cm |
---|
Resistivity Resolution | dependent on resistivity reading |
---|
Resistivity Calibration | based on conductivity calibration |
---|
Salinity Range | 0.00 to 70.00 PSU |
---|
Salinity Resolution | 0.01 PSU |
---|
Salinity Accuracy | ±2% of reading or ±0.01 PSU whichever is greater |
---|
Salinity Calibration | based on conductivity calibration |
---|
Seawater σ Range | 0.0 to 50.0 σt, σ₀, σ₁₅ |
---|
Seawater σ Resolution | 0.1 σt, σ₀, σ₁₅ |
---|
Seawater σ Accuracy | ±1 σt, σ₀, σ₁₅ |
---|
Seawater σ Calibration | based on conductivity calibration |
---|
DO Range | 0.0 to 500.0%; 0.00 to 50.00 ppm (mg/L) |
---|
DO Resolution | 0.1%; 0.01 ppm (mg/L) |
---|
DO Accuracy | 0.0 to 300.0% saturation: ±1.5% of reading or ±1.0% saturation whichever is greater, 0.00 to 30.00 ppm (mg/L): ±1.5% of reading or ±0.10 ppm (mg/L), whichever is greater, 30.00 ppm (mg/L) to 50.00 ppm (mg/L): ±3% of reading, 300.0 to 500.0% saturation: ±3% of reading |
---|
DO Calibration | automatic one or two points at 0 and 100% or one custom point |
---|
Atmospheric Pressure Range | 450 to 850 mm Hg; 17.72 to 33.46 in Hg; 600.0 to 1133.2 mbar; 8.702 to 16.436 psi; 0.5921 to 1.1184 atm; 60.00 to 113.32 kPa |
---|
Atmospheric Pressure Resolution | 0.1 mm Hg; 0.01 in Hg; 0.1 mbar; 0.001 psi; 0.0001 atm; 0.01 kPa |
---|
Atmospheric Pressure Accuracy | ±3 mm Hg within ±15°C from the temperature during calibration |
---|
Atmospheric Pressure Calibration | automatic at one custom point |
---|
Temperature Range | -5.00 to 55.00 °C, 23.00 to 131.00 °F, 268.15 to 328.15 K |
---|
Temperature Resolution | 0.01 K, 0.01 °C, 0.01 °F |
---|
Temperature Accuracy | ±0.15 °C; ±0.27 °F; ±0.15 K |
---|
Temperature Calibration | automatic at one custom point |
---|
Temperature Compensation | automatic from -5 to 55 °C (23 to 131 °F) |
---|
Logging Memory | 45,000 records (continuous logging or log-on-demand of all parameters) |
---|
Logging Interval | one second to three hours |
---|
PC Connectivity | USB to PC with HI9298194 Hanna software installed |
---|
Battery Type/Life | 1.5V AA batteries (4) / approximately 360 hours of continuous use without backlight (50 hours with backlight) |
---|
Environment | 0 to 50°C (32 to 122°F); RH 100% (IP67) |
---|
Dimensions | 185 x 93 x 35.2 mm (7.3 x 3.6 x 1.4”) |
---|
Weight | 400 g (14.2 oz.) |
---|
Ordering Information | HI98194 is supplied with HI7698194 probe, HI7698194-1 pH/ORP sensor, HI7698194-3 EC sensor, HI7698194-2 DO sensor, HI 76981942 Probe maintenance kit (HI 7042S DO electrolyte solution, DO membrane caps (5), O-rings for sensors (5), syringe with grease to lubricate the O-rings), HI9298194 Hanna PC software, HI920015 micro USB cable, 1.5V AA batteries (4), instruction manual, quick start guide, quality certificate, and rugged carrying case with custom thermoformed insert. |
---|
Ulasan